The ASAP 2020 is an adsorption desorption analyzer for measuring surface area, pore size, and pore volume of powders and porous materials. Standard methods or user customized protocols can be used to characterize adsorbents, catalysts, zeolites, MOFs, APIs, excipients, and a wide variety of porous and non-porous materials.
The ASAP 2020 uses two independent vacuum systems, one for sample analysis and one for sample preparation. This allows preparation and analysis to proceed concurrently without the inherent delay found in single vacuum system analyzer that must share a pump. Moreover, independent systems completely eliminate the possibility of cross-contamination between the degas and analysis manifolds
Surface area method Relative pressure (P/P0) – from 0.0500 to 0.3000 Data obtained • BET SURFACE AREA (m2/g) • LANGMUIR SURFACE AREA (m2/g) MESOPORE WITH BJH - (degassing 1 step) Relative pressure (P/P0) – from 0.060000 to 0.989000 (BJH ADS), from 0.989000 to 0.24000 (BJH DES), Data obtained • BET SURFACE AREA (m2/g) • LANGMUIR SURFACE AREA (m2/g) • t-PLOT • BJH ADSORPTION • BJH DESORPTION • TOTAL PORE VOLUME (cm3/g) MICROPORE ANALISIS TOTAL 60 POINT (degassing 2 step) Relative pressure (P/P0) – from 0.010000 to 0.989000 (BJH ADS), from 0.989000 to 0.24000 (BJH DES), Data obtained • BET SURFACE AREA (m2/g) • LANGMUIR SURFACE AREA (m2/g) • t-PLOT • BJH ADSORPTION • BJH DESORPTION • HORVATH-KAWAZOE • DUBININ-ASTAKHOR • TOTAL PORE VOLUME (cm3/g)